Artigo
Development of an electrochemical sensor modified with dealuminated zeolite with citric acid for hydroxyzine determination by BIA-Amperometry
Carregando...
Notas
Data
Orientadores
Editores
Coorientadores
Membros de banca
Título da Revista
ISSN da Revista
Título de Volume
Editor
Visão Fokka Communication Agency
Faculdade, Instituto ou Escola
Departamento
Programa de Pós-Graduação
Agência de fomento
Tipo de impacto
Áreas Temáticas da Extenção
Objetivos de Desenvolvimento Sustentável
Dados abertos
Resumo
Abstract
An electrochemical sensor modified with
zeolite dealuminated with citric acid was
developed for the determination of
hydroxyzine in pharmaceutical products
during BIA-amperometry experiments. The
modified electrochemical sensor was
prepared by mixing powdered graphite with
zeolite treated with citric acid homogenized
with hexane and mineral oil (60:20:20%
w/w/w respectively). The developed sensor
showed reproducible amperometric responses in a linear range of 1.0x10-6 – 2.0x10-5 mol L-1 in +1,28 V
(vs. Ag/AgCl). The detection limit and detection limit found were 3.10x10-7 mol L-1 e 1.04x10-6 mol L-1,
respectively. The method developed by BIA-amperometry was applied for the determination of the
analyte in tablets and commercial syrups and the results found agreement with the nominal values of the
commercial samples.